Deleting reagents
Highlight the reagent to be deleted in the SET UP REAGENTS AND
WARNING THRESHOLDS screen.
Press DELETE.
Press OK in the screen prompt.
Please remember that a reagent which is already used in a pro-
gram cannot be deleted.
Adding new reagents to stations
Go to SMART FUNCTIONS to remote-fill the reagent bottle.
If a bottle is filled manually, it must be defined as full afterwards.
For that purpose:
Press MENU to access the MENU FUNCTIONS screen.
Press REAGENT STATUS.
In the reagent list, highlight the station / reagent that has been filled
manually.
Press SET AS FULL to mark the station as ‘full’.
Upon activation of the RMS all warning thresholds for the se-
lected reagent are automatically reset to ‘0’.
5. Operation